TU5200 Laboratory Laser Turbidimeter EPA Version

Only the new TU5 Series Lab & Process Turbidimeters with 360° x 90° Detection deliver unprecedented confidence that a change in your reading is a change in your water.

SKU: TU5200 Laboratory Laser Turbidimeter without RFID, EPA Version Category: Tags: ,

TU5 Series® TU5200 Laboratory Laser Turbidimeter without RFID, EPA Version

The next standard in the evolution of turbidity

The platform employs a patented optical design that sees more of your sample than any other turbidimeter, delivering the best low level precision and sensitivity while minimizing variability from test to test. For the first time you will be able to remove the uncertainty of which measurement to trust, thanks to identical 360° x 90° Detection® Technology in both, lab and process, instruments. The TU5 Series® dramatically reduces the time needed to get a turbidity measurement you can rely on.
The TU5200 Lab Turbidimeter, with the 360° x 90° Detection Technology, sees more of your sample meaning you no longer have to apply silicone oil to your vial and index to ensure an accurate reading. In addition, you can have confidence that you will get consistent results in the lab, from operator to operator, since the impact of small interferences is minimized due to the collection of light at 360°.

Only the new TU5 Series Lab & Process Turbidimeters with 360° x 90° Detection deliver unprecedented confidence that a change in your reading is a change in your water.

HACH

Accuracy

±2 % plus 0.01 NTU from 0 – 40 NTU;
±10 % of reading from 40 – 1000 NTU based on Formazin primary standard (at 25 °C)

Calibration Method

For Formazin and Stablcal:
20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range

Custom calibration for up to 6-point calibrations

Degrees 0 – 100 mg/L
SDVB 0 – 1000 NTU

Certifications

CE compliant
US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version
Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Australian ACMA Marking

Communication

Link2sc

Dimensions (H x W x D)

195 mm x 409 mm x 278 mm

Display

7" WVGA (800 x 480) color TFT-LCD with touch screen

Enclosure Construction

PC/ABS Blend Plastic

Light Source

Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)

Model

TU5200 EPA

Operating Humidity

80% at 30 °C (non condensing)

Operating Temperature Range

10 – 40 °C (50 – 104 °F)

Power Requirements (Hz)

50/60 Hz

Power Requirements (Voltage)

100 – 240 VAC

Range

EPA:
0 – 700 NTU / FNU / TE/F / FTU
0 – 100 mg/L
0 – 175 EBC

ISO:
0 – 1000 NTU / FNU / TE/F / FTU
0 – 100 mg/L
0 – 250 EBC

Regulatory

EPA

Repeatability

<40 NTU: Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater
>40 NTU: Better than 3.5% of reading on Formazin at 25 °C (77 °F)

Resolution

0.0001 NTU / FNU / TE/F / FTU / EBC / mg/L

Response Time

Signal averaging off: 7 seconds
Signal averaging on: 10 seconds (when averaging time is 5 seconds)

Sample Temperature

4 – 70 °C (39 – 158 °F)

Source Lamp

EPA Version – 650 nm Class 2 Laser

Storage Conditions

-30 – 60 °C (-22 – 140 °F)

Stray Light

<10 mNTU

Units

NTU, FNU, TE/F, FTU, EBC;

mg/L if calibrated with Degrees calibration curve

Warranty

12 months

Weight

2.4 kg (5.29 lbs.)

What's included?

TU5200 benchtop turbidimeter, user manual, Stablcal kit (10, 20 and 600 NTU), 2 sample vials, power supply, dust cover, vial stand

brand

هیچ ویدیویی برای این محصول موجود نیست.

Groundbreaking 360° x 90° Detection Technology

The TU5 Series employs a patented optical design that sees more of your sample than any other turbidimeter, delivering the best low level precision and sensitivity while minimizing variability from test to test.

Everything about turbidity – faster

The TU5 Series dramatically reduces the time needed to get a turbidity measurement you can rely on, with 98% less online sample surface area to clean, sealed vials for calibration, and the elimination of the need for indexing and silicone oil in the lab. Not to mention, a smaller online sample volume means you will detect events almost immediately.

Matching lab and online results

For the first time you will be able to remove the uncertainty of which measurement to trust, thanks to identical 360° x 90° Detection™ Technology in both instruments.