Grease penetration tester LinteL PN-10 (equipment C)
Apparatus for determining the penetration of greases
SKU:
LinteL PN-10
Category: laboratory equipment
- Implemented standards: GOST 5346-78, GOST ISO 2137
- LinteL PN-10 allows you to set the following test conditions:
– penetration time in increments of 1s;
– delay time before penetration is performed in steps of 1s.
- The device determines the penetration value with an error of no more than ±0.05mm
- The device is equipped with a built-in illuminator that helps determine the moment the cone touches the product being tested
- The device provides storage of up to 9 determination results with calculation of their average value
- Integration with the LinteL Link data collection system allows you to collect and transfer laboratory test results from the device to a personal computer via wireless communication. Integration with the laboratory information system Lintel LIS provides comprehensive automation of laboratory activities
- Control of table movement in three modes: slow table lift, fast table lift, fast table return
- Display on the display in standby mode: the number of the last test performed; a given number of tests; table lifting speed; specified penetration time; specified delay time
- Display on the display during the test: the number of the last test performed; a given number of tests; penetration time (countdown); current penetration value
- The four-line OLED display provides an intuitive user interface with all the information you need while testing and viewing results.
- Full self-diagnosis system with indication of the causes of malfunctions on the display
- Stopping the test process at the end of the analysis with an audible alarm
Safety |
|
Sound accompaniment |
Sound signal when the test is completed or a malfunction is detected |
Diagnostics and setup |
Built-in self-diagnosis and configuration algorithms, notifying users about the causes of malfunctions |
Penetration measuring range |
0 to 630 penetration units |
---|---|
Displacement meter error |
no more than ± 0.1 mm |
Penetration time |
from 1 to 3599 s |
Delay time before penetration |
from 0 to 3599 s |
Table speed |
from 0.02 to 5 mm/s |
Power consumption |
no more than 50 W |
Dimensions |
280 x 245 x 505 mm, (depth x width x height) |
Weight |
no more than 12 kg |
Ambient temperature |
from +15 to +35 °C |
Relative humidity |
from 75% |
Voltage |
from 187 to 242 V |
Frequency |
50±1 Hz |
Life time |
6 years, no more than 15,000 hours |
Warranty period |
1 year, no more than 2,500 hours |
Standards | |
brand |