Automatic Flash point closed cup tester Tag Lintel ATVT-20
full automation of the testing process of flash point
SKU:
ATVT-20
Category: laboratory equipment
- The Lintel ATVT-20 device provides full automation of the testing process – filling the bath with water, maintaining the heating rate of the product, testing the product with built-in electric spark ignition, determining and storing the flash point with automatic adjustment to atmospheric pressure, replacing a heated coolant with a cold one
- Automatic coolant renewal after the end of the test reduces the time between tests
- Integration with the LinteL Link data collection system allows you to collect and transfer laboratory test results from the device to a personal computer via wireless communication. Integration with the laboratory information system Lintel LIS provides comprehensive automation of laboratory activities
- High-contrast color TFT display provides an intuitive user interface with all the information you need while testing and viewing results
- Test log maintenance provides storage in non-volatile memory of up to 100 test results (test method, product name, expected flash point, flash point, atmospheric pressure, heating rate, date and time of test completion, performer), as well as a graph of heating rate during the last test
- Automatic stop of the test with a sound signal about the end of the test
- Detection of product overheating with automatic end of test
- High precision product temperature sensor
- Full self-diagnosis system with indication of the causes of malfunctions on the display
EQUIPMENT:
1.LinteL ATVT-20 device – 1 pc.;
- Crucible cover – 1 pc.;
- Crucible – 2 pcs.;
- Screw clamp 12-20 – 4 pcs.;
- Reinforced tube 1/2″ 1 m – 2 pcs.;
- Pressure reduction valve Honeywell D04FS-1/2″ A – 1 pc.;
- Operation manual – 1 pc.;
- Passport – 1 pc.
Flash point detection range |
from +13 to +93 °C |
---|---|
Product temperature measurement resolution |
0.1 °C |
Product heating rate, depending on the selected test conditions |
1±0.1°C/min, 3±0.3 °C/min |
Atmosphere pressure |
from 84 to 106.6 kPa (from 630 to 800 mm Hg) |
Refrigerant temperature (water) |
from + 4 to +15 °C |
Refrigerant flow (water) |
from 2.5 to 5.0 l/min |
Heating rate for products with a flash point below plus 60°C |
from 0.9 to 1.1 °C/min |
Heating speed for products with a flash point of plus 60°C and above |
from 2.7 to 3.3 °C/min |
Atmospheric pressure sensor error |
mmHg. ±(12.0+α)* |
Power consumption |
no more than 500 VA |
Device dimensions |
370 x 420 x 310 mm (depth x width x height) |
Device weight |
no more than 14 kg |
Ambient temperature |
from +15 to +35 °C |
Relative humidity |
no more than 75% |
Mains voltage |
from 187 to 242 V |
Power frequency |
49 to 51 Hz |
Life time |
6 years, no more than 15,000 hours |
Warranty period |
1 year, no more than 2,500 hours |
Standards | |
brand |