Automated Pensky Martens flash point analyzer (closed cup) LinteL ® ATV-21

flash point determination with an electric spark ignition

SKU: pensky-martens Category:
  • LinteL АTV-21 analyzer provide complete automation of the testing process. Pre-installed test programs. Each test program contains test standard and parameters. When choosing a program, all the necessary parameters and test conditions are set and maintained automatically.
  • Unique design of built-in electric spark ignition – there is no open flame, there is no risk of fire in the laboratory. Accuracy of flash point determination with an electric spark ignition does not differ from testers with gas ignition.
  • Test logging provides storage of up to 1,000 test results and the last 100 diagrams.
  • Automatic mixing of the sample and programmed increase of the product temperature at a specified rate.
  • Possibility of product preheating switching on to reduce test time for products with a high flash point.
  • Calculation of average flash point.
  • Convenient intuitive menu for selecting the analysis program, navigation in the test log, diagnostics and setup of the tester.
  • Automatic correction for barometric pressure in the ranges from 630 to 810 mm Hg.
  • Complete self-diagnosis system with fault indication on the display.
  • Test stop at the end of analysis with audible alarm.
  • Color touchscreen LCD is used as a data input (keyboard) and output device simultaneously.
  • Analyzer can be used for quality control of vegetable oils.
  • Test by “Express” method for products with unknown flash point.
  • Integration with LinteL Link Data Collection System enables collection and transfer of laboratory test results from the tester to a personal computer via wireless communication. Integration with the laboratory information management system LinteL LIMS provides complex automation of laboratory activities.
SAFETY
Electrical safety Differential automatic circuit breaker, preventing electric shock for the user and the tester damage due to short circuit
Fire safety When readings on the digital temperature meter exceed plus 395 °С, or 28 spark tests have been completed, the test is stopped. Tester issues an intermittent audible alarm and displays the error message on the screen. When the test is stopped, the heating is turned off.
Audio signals Beeping at the end of the test and failure detection
Diagnostic and setting Built-in algorithms for self-diagnostics and settings, user notification about causes of malfunctions
Flash point range

Automatic flash point determination in temperature ranges from +12 up to +370 °C

Heating

Automatic product heating and maintaining the product heating rate according to the selected test program

Ignition

Automatic ignition with an electric spark of controlled capacity for complete simulation of gas ball 4 mm in diameter

Stirrer

Stirring rate corresponds with selected test program requirements and is maintained automatically

Power supply

750 W, max:

Dimensions (with cup installed)

420 х 310 х 455 mm (W x H x D)

Weight

15 kg max

Ambient air temperature

10 to 35 °С

Relative air humidity at a temperature of +25°С

80 % max

Voltage

198 to 242 V

Frequency

50 ± 1 Hz

Service life

6 year (15,000 hours max)

Warranty period

1 year (2,500 hours max)

Standards

,

brand

ASTM D93 ISO 2719