Automatic open cup tester (with gas ignition) Cleveland method LinteL®ATVO-21
measure the flash point temperature and flash point of all petroleum and hydrocarbon products and cuts
This test method is used to measure the flash point temperature and flash point of all petroleum and hydrocarbon products and cuts that have a flash point of at least 79 and a maximum of 380 degrees Celsius.
The Clevland open-cup method is used to measure the open flash point of all petroleum products with an open flash point in the temperature range of at least 79 and at a maximum of 400 ° C, except for furnace oil.
Safety
Beeping at the end of the test and failure detection |
Audio signals |
Built-in algorithms for self-diagnostics, user notification about causes of malfunctions |
Diagnostic and setting |
Differential automatic circuit breaker, preventing electric shock for the user and the tester damage due to short circuit, complete tester disconnection from power supply under all modes, except test one. |
Electrical safety |
Measuring range |
from +79 to +400 °С |
---|---|
Discreteness of the flash-point result determination |
1 °C |
Type of the product vapor ignition |
gas ignition |
Product heating rate |
5–6 °C/min |
Product heating rate up to temperature by 56 °C below expected flash point |
14–17 °C/min |
ACCURACY PARAMETERS |
5 °C max |
Reading repeatability limit of the tester |
5 °C max |
Reading reproducibility limit |
16 °С max |
Barometric pressure |
±1 mm Hg |
Power consumption under standby mode |
150 max |
Power consumption under test mode |
1,000 W max |
Overall dimensions |
330 х 420 х 310 mm (depth x width x height) |
Weight |
21 kg max |
Ambient air temperature |
from +15 to +35 °С |
Relative air humidity |
75 % max |
Voltage |
from 187 to 242 V |
Frequency |
from 49 to 51 Hz |
Probability of failure-free operation within 1,000 h, min |
0.96 |
Service |
6 year, but 15,000 hours max |
Guaranteed service life |
1 year, but 2,500 hours max |
Standards | |
brand |