Catalytic filters

ntended for gas cleaning from oxygen and organic impurities by heating in the presence of catalyst.

Intended for gas cleaning from oxygen and organic impurities by heating in the presence of catalyst. Catalytic filter is commonly used for nitrogen cleaning from oxygen when working with ECD as well as with WAX or FFAP column types, and aslo for cleaning of compressed air from organic impuryties when feeding flame detectors, or when air is used as carrier gas. Oxygen absorption channel can be recovered with hydrogen flow.

Chromatec supplies 3 models of catalytic Filter 20.0:

  • Filter 20.0-01 – for air cleaning from organics (up to 0.1 mg/m3 and less)
  • Filter 20.0-02 – for carrier gas cleaning from oxygen (up to 5 ppmv and less)
  • Filter 20.0-03 – dual-channel filter combining -01 and -02 models
Operating temperature of reactor, ˚С

400

Warm-up time, min, not more

30

Maximum air flow, ml/min

700

Max flow of inert gas, ml/min

500

Maximum content of organics in output air, as a methane, mg/m3

0/1

Oxygen trapping capacity, L

10

Maximum pressure drop at reactor

0/08

Dimensions: (WxDxH);

219 mm x 168 mm x 327 mm

Fittings

1/8

Storage Temperature:

from -50 °C to 50°C