Bitumen Direct Tensile Testing Machine DB–20–100
Lintel DB-20-100 provides automatic testing of bitumen samples for ductility, elasticity, deformation energy, as well as testing of plastics and polymers for tension.
Lintel DB-20-100 provides automatic testing of bitumen samples for ductility, elasticity, deformation energy, as well as testing of plastics and polymers for tension
The device provides a specified temperature in the test zone due to the built-in thermo cryostat from -10 to +40°C
- Lintel DB-20-100 allows testing from 1 to 3 samples with a maximum breaking force of up to 30 kg per sample
- Automatic movement of the carriage at a given speed
- Integration with the Lintel Link data collection system allows for collection and transmission of laboratory test results from the device to a personal computer via wireless communication. Integration with the Lintel LIS laboratory information system provides comprehensive automation of laboratory activities.
- A built-in pump for mixing the bathhouse coolant and a modern cooling system ensure a minimum temperature gradient in different parts of the bathhouse
- Test logging provides storage of up to 100 test results and 30 sample loading graphs
- Memorization of the extensibility value and sample number in automatic mode by the drop in tension force or by the operator pressing the remote button for fixing the sample break
- Possibility of checking the temperature sensor readings of the bath with a standard thermometer
- The operator interaction system based on a color touch screen facilitates the learning and daily operation of the device
- The set of disassemblable forms for samples supplied with the device comply with GOST, ASTM, EN and provide convenient preparation of samples for testing.
- Transparent ergonomic lid
- Full self-diagnostic system with indication of causes of malfunctions on the display
- Stopping the testing process at the end of the analysis with an audible signal
- Possibility of starting the thermostat by timer – “delayed start”
Safety
Electrical safety: Grounding of non-current-carrying parts and compliance with protection class 0I according to GOST R IEC 61140-2000
Sound accompaniment: Sound signal at the end of the test, detection of a fault
Diagnostics and setup: Built-in self-diagnostics and configuration algorithms, notifying users about the causes of malfunctions
Maximum length of elongation of test samples |
1,000 mm |
---|---|
The number of simultaneously tested samples |
from 1 to 3 |
Maximum load on each sample |
300 N |
Carriage speed |
from 0.1 to 100 cm / min |
Bath volume |
17 l |
Range of maintained bath temperature |
from -10 to +40 ° C |
Force measurement accuracy |
± 1% |
Irregularity of the temperature of the coolant in the bath |
not more than 0.5 ° C |
Temperature measurement error |
no more than ± 0.2% |
Carriage Speed Accuracy |
no more than ± 0.25 cm / min |
Length measurement accuracy: – in the range from 0 to 300mm |
no more than ± 2 mm |
Length measurement accuracy: – in the range from 300 to 1 000mm |
no more than ± 3 mm |
Shape deviation |
meet standards |
Power consumption |
no more than 650 W |
Dimensions |
370 x 260 x 1750 mm, (depth x height x width) |
Weight |
no more than 50 kg |
Ambient temperature |
from +15 to +35 ° C |
Relative humidity |
humidity at a temperature of + 20 ° С, not more than 75% |
Voltage |
from 187 to 242 V |
Warranty period of operation |
1 year, no more than 2 500 hours |
Standards |
ASTM D 113-99, ASTM D 6084-04, EN 13398: 2010, EN 13587: 2010, EN 13589: 2008, EN 13703: 2003 |