Wavelength dispersive X-Ray fluorescence analyzer for sulfur and chlorine in liquids quantification.

SKU: spectroscan CLSW Category:

SPECTROSCAN CLSW is a reliable and accurate WDXRF analyzer of sulfur and chlorine in liquids. It provides quick determination of sulfur and chlorine weight content with Limit of Detection as low as 0.3 ppm for sulfur and 0.2 ppm for chlorine.

SPECTROSCAN CLSW fully complies with international standards for sulfur and chlorine analysis by WDXRF technique.
Advanced design, intuitive menu and extended set of CRMs and consumables in starter pack makes easy the work even for a non-professional analyst.

Advantage:

Due to the original design no He blowdown is required, providing LOD as low as 0.2 ppm for CI and 0.3 ppm for S

  • Simple in operation;
  • Desk top instrument;
  • The sample data and analysis results are shown in display and printed out by inbuilt printer;
  • Special ventilated cups are developed for volatile samples;
  • 3-position sample automatic changer makes analysis of sample replicateseasy.

Due to lateral position of the sample:

  • Error due to water and air bulbs excluded;
  • Contamination of the inner parts of the analyzer is excluded;
  • Additional errors due to contamination of the additional protection film is excluded;
  • Sample loading unit is easily cleaned.

Minimum operator actions:

  • Sample preparation;
  • Put in the number or name of the sample from inbuilt key-board; 
  • Fill in two sample cups with the sample;
  • Put in the samples into the analyzer and start measurements.

All the rest operations are run automatically.

Application:

SPECTROSCAN CLSW provides determination of chlorine and sulfur weight content in all types of liquids with Limits of Detection as low as 0.2 ppm for chlorine and 0.3 ppm for sulfur.

SPECTROSCAN CLSW fully complies with international standards for chlorine and sulfur analysis by WDXRF technique.

Measurement principle

Energy dispersive X-ray fluorescence (EDXRF)

Sulfur and chlorine analytical lines selection

Crystal diffraction

Crystal analyzer

Doubly-curved pyrolytic graphite (C, 002)

Sample compartment environment

Ambient air

Optical chamber environment

Vacuum, No Helium blowdown needed

Detector

Sealed proportional counter

Measurement time

10 – 999 seconds user selectable. Typical time for one sample – 200 seconds

Sample changer

Automatic lateral loading for 3 sample cups

Sample cups, diameter, volume

∅32 mm, V 8 cm3, ∅32 mm, V 8 cm3, ventilated

X-ray tube capacity

less 200 W

User interface

display, thermo printer

Results print-out

Built-in thermal printer

Outputs

RJ-45, RS-232, USB

Dimensions

530х480х340 mm – spectrometer, 330х230х380 mm – vacuumpump

weight (not more)

40 kg– spectrometer, 9 kg – vacuumpump

Power supply

220 V, ~ 50 Hz, 750 W

Standards

, , ,

brand

ASTM D4929 ASTM D7039 ASTM D7536 NB/SH/T 0842